Axios & Zetium System Inspector

Drop in your SuperQ Userdata folder to check X-ray tube output, flow & scintillation detector drift, crystal alignment, and hardware error logs right in your browser.

Using a benchtop Epsilon 1, 3, or 4? Open the Epsilon System Inspector →
Axios and Zetium WDXRF System Inspector icon
Drag & Drop Your C:\PANalytical\SuperQ\Userdata Folder Here
Supports Axios, Axios mAX, Axios FAST, Zetium, MagiX, and PW2400 / PW2404 / PW2440 systems running SuperQ
๐Ÿ”’ 100% Private Browser Processing: Your files stay on your PCโ€”nothing is uploaded to our server, and exported .7z packages are AES-256 encrypted locally in your browser.
Important Tip for Every SuperQ User

Turn On SuperQ’s Built-In Sensor Logger (Takes 60 Seconds)

SuperQ can continuously log your spectrometer’s vacuum, water flow, water conductivity, cabinet temperature, and P10 gasโ€”but PANalytical ships this setting turned OFF by default. Turning it on makes future troubleshooting dramatically faster.

How to Turn It On Inside SuperQ

On your SuperQ Manager launcher bar, click the System Setup icon (the spectrometer cabinet in front of the colorful periodic table, next to Measure and Analyse).
  1. Open SuperQ Manager and click the System Setup icon.
  2. In the top menu bar, click System → System log configuration...
  3. Check Enable system log at the top.
  4. Check all 9 parameters (Vacuum level, Cabinet temperature, Conductivity, Internal flow, External flow, Primary temperature, Gas flow, Gas pressure, Medium pressure).
  5. Set Log every 01:00 (hourly), check When measurement starts, and click OK.

Why This Matters for Future Troubleshooting

  • Cooling Water & Chiller Protection: Tracks internal/external flow (L/min) and DI water conductivity (µS/m) so clogged strainers, worn chiller pumps, and spent DI filters show up weeks before a generator trip.
  • Vacuum Leak Detection: Records chamber vacuum (Pa) on every run so light-element drift (Na, Mg, Al, Si) is immediately traced to an airlock seal or backing pump instead of chasing calibrations.
  • Temperature Drift Checks: Logs cabinet and water temperatures (°C) to spot clogged dust filters or overnight room HVAC swings shifting crystal angles.
  • P10 Gas Stability: Logs detector gas flow (L/h) and pressure (hPa) so empty bottles or regulator drift are obvious right away.
Need on-site repair, PM, or tested replacement parts for an Axios, Zetium, or MagiX? Call 630.229.3859.